Reset to Factory Defaults
>>> Tools: Reset to Factory Default
>>> Ctrl+F
The Reset to Factory Default function is used to reset all configuration and calibration settings within the logger.
WARNING - This will cause the loss of any calibration values and all data that may be stored in the logger.
Some Logger Types have more than one Factory default.
For example the pH-HR pH logger has two factory defaults - one for a probe that has a PT100 temperature sensor in the pH probe and one for a probe that has no temperature sensor and therefore has no temperature compensation of pH.
The GP-HR General Purpose Logger has a factory default for each of the available standard probesets.
If a logger has more the one factory default, a list of the available factory defaults will be displayed in the Reset Logger to Factory Default dialog box.

Select the default required from the list provided

For example if you wish to set up a GP-HR General Purpose Logger as a 2 channel 4 to 20 mA logger, select GP-HR mA2.
Then click on the Reset (Logger Name) to Factory Default button.
The following list shows loggers that have more than one factory default and a discription of the factory default codes used in the OmniLog Software.
SafeT
- SafeT Int Temp Only
- Only the Internal Temperature Sensor is used. The External channel is set to "None"
- SafeT with Ext Temp
- Channel 1 is Internal Temperature Sensor
- Channel 2 is External Temperature Probe (Thermister)
GP-HR General Purpose High Resolution Datalogger
- GP-HR General
- Used for testing all inputs
- Channel 1 to 3 are set to analog to digital converter numbers
- Channel 4 is set to counter
- GP-HR R-PS 0.1mm
- Rain Probe Set
- Channel 1 set to Internal Temperature
- Channel 2 to 3 are set to "None"
- Channel 4 is set for 0.1mm Tipping Bucket Rain Gauge
- GP-HR R-PS 0.2mm
- Rain Probe Set
- Channel 1 set to Internal Temperature
- Channel 2 to 3 are set to "None"
- Channel 4 is set for 0.2mm Tipping Bucket Rain Gauge
- GP-HR R-PS 1mm
- Rain Probe Set
- Channel 1 set to Internal Temperature
- Channel 2 to 3 are set to "None"
- Channel 4 is set for 1mm Tipping Bucket Rain Gauge
- GP-HR RT-PS 0.1mm
- Rain Probe Set
- Channel 1 set to External Temperature Probe (Thermister)
- Channel 2 set to Internal Temperature
- Channel 3 set to "None"
- Channel 4 is set for 0.1mm Tipping Bucket Rain Gauge
- GP-HR RT-PS 0.2mm
- Rain Probe Set
- Channel 1 set to External Temperature Probe (Thermister)
- Channel 2 set to Internal Temperature
- Channel 3 set to "None"
- Channel 4 is set for 0.2mm Tipping Bucket Rain Gauge
- GP-HR RT-PS 1mm
- Rain Probe Set
- Channel 1 set to External Temperature Probe (Thermister)
- Channel 2 set to Internal Temperature
- Channel 3 set to "None"
- Channel 4 is set for 1mm Tipping Bucket Rain Gauge
- GP-HR T1-PS
- Single Temperature Probe Set
- Channel 1 set to External Temperature Probe (Thermister)
- Channel 2 and 3 set to "None"
- Channel 4 set to Internal Temperature
- GP-HR T2-PS
- Dual Temperature Probe Set
- Channel 1 and 2 set to External Temperature Probe (Thermister)
- Channel 3 set to "None"
- Channel 4 set to Internal Temperature
- GP-HR T3-PS
- Three Temperature Probe Set
- Channel 1 to 3 set to External Temperature Probe (Thermister)
- Channel 4 set to Internal Temperature
- GP-HR mA1
- Single input milliAmp adapter
- Channel 1 set to milliAmp
- Channel 2 and 3 set to "None"
- Channel 4 set to Internal Temperature
- GP-HR mA2
- Dual input milliAmp adapter
- Channel 1 and 2 set to milliAmp
- Channel 3 set to "None"
- Channel 4 set to Internal Temperature
- GP-HR mA3
- Three input milliAmp adapter
- Channel 1 set to milliAmp
- Channel 2 and 3 set to "None"
- Channel 4 set to Internal Temperature
- GP-HR mA2 ITemp
- Dual input milliAmp adapter
- Channel 1 and 2 set to milliAmp
- Channel 3 set to Internal Temperature
- Channel 4 set to set to "None"
- GP-HR Pt1000 mA 5V
- Channel 1 set to Pt1000 Temperature Probe
- Channel 2 set to milliAmp input
- Channel 3 set to 5Volt input
- Channel 4 set to Internal Temperature
ph-HR
- ph-HR without Temp
- pH with no temperature sensor in the pH probe
- Channel 1 set to Internal Temperature
- Channel 2 set to pH
- Channel 3 and 4 set to "None"
- pH-HR Temp Comp
- pH with a Pt100 sensor in the pH probe - this sensor is used for temperature compensation of the pH values.
- Channel 1 set to Pt100 sensor in the pH probe
- Channel 2 set to pH
- Channel 3 set to Internal Temperature
- Channel 4 set to "None"