Leaf Wetness Calibration
The Leaf Wetness sensor is calibrated to read 0%Leaf Wetness when the sensor surface is dry and 100% when the surface is very wet.
- Dry the Leaf Wetness sensor
- Enter 0 into the First Point Actual Value box
- Click on the First Point button
- Sprinkle water over the Leaf Wetness sensor.
- Enter 100 into the Second Point Actual Value box
- Click on the Second Point button
- Click on the Write Unit, Scaling and Calibration Values to logger button
The Calibration process is now complete