Tipping Bucket Rain Gauge Calibration
- Enter zero in the Calibration Offset box
- Enter the Volume per Tip value as supplied by the manufacture of the rain gauge, into the Calibration Gain box. ~ for example:
Volume per Tip
|
Gain
|
0.1mm
|
0.1
|
0.2mm
|
0.2
|
0.5mm
|
0.5
|
1mm
|
1
|
2mm
|
2
|
- These Offset and Gain figures should now be written to the logger
- Click on the Write Unit, Scaling and Calibration values to the Logger button
To check the calibration
- Stop the logger, if it is running
- Select the Realtime screen
- Manually tip the bucket on the rain gauge 10 times
- Check that each tip shows as an increase in rain fall on the realtime screen
By default the Realtime screen updates every 2 seconds but this can be changed from the Options screen. You will have to wait for a Realtime screen update before the new rainfall values show on the rainfall gauge.