J Thermocouple Calibration
J Thermocouple Calibration Procedure using a calibration voltage source
- Connect the calibration voltage source to the logger observing the correct polarity
- Set the calibration source to 0.0 mV
- Enter 0 in to the box labeled First Point Actual Value
- Note: If you are using °F enter 32 instead of 0
- Click the First Point Button
- Set the calibration source to 57.94 mV
- Enter 1000 in to the box labeled Second Point Actual Value
- Note: If you are using °F enter 1832 instead of 1000
- Click the Second Point Button
- The Calibration Offset and Calibration Gain figures will now be calculated and shown on the calibration screen.
- These Offset and Gain figures should now be written to the logger
- Click the Write Unit, Scaling and Calibration Values to logger button
- The Calibration process is now complete
To check the Calibration
- Go to the Realtime screen
- Set the calibration source to 0.0 mV
- The J Thermocouple gauge on the Realtime screen should show the ambient temperature (Ambient Temperature plus 0°C)
- Set the calibration source to 57.94 mV
- The J Thermocouple gauge on the Realtime screen should show Ambient Temperature plus 1000°C
This calibration procedure is based on the following data.
- a J thermocouple junction gives 0.0mV when the thermocouple and the logger are at the same temperature.
- It will give 57.94mV when there is a 1000degC difference between the logger and the thermocouple.
J Thermocouple Calibration Procedure °C using two known temperatures
- Connect the J thermocouple sensor to the logger
- Allow the J thermocouple sensor to stabilize at the first known (lower) temperature
- Read the Ambient Temperature of the Logger using the Realtime Screen
- Enter the first known temperature MINUS THE AMBIENT TEMPERATURE in to the box labeled First Point Actual Value
- Click the First Point Button
- Allow the J thermocouple sensor to stabilize at the second known temperature
- Read the Ambient Temperature of the Logger using the Realtime Screen
- Enter the second known temperature MINUS THE AMBIENT TEMPERATURE in to the box labeled Second Point Actual Value
- Click the Second Point Button
- The Calibration Offset and Calibration Gain figures will now be calculated and shown on the calibration screen.
- These Offset and Gain figures should now be written to the logger
- Click the Write Unit, Scaling and Calibration Values to logger button
- The Calibration process is now complete
If this procedure is used the J Thermocouple will show the correct temperature when sitting in free air.
If the ambient temperature is not subtracted from the known temperatures the J thermocouple will read high by the amount of the ambient temperature.
J Thermocouple Calibration Procedure °F using two known temperatures
- Connect the J thermocouple sensor to the logger
- Allow the J thermocouple sensor to stabilize at the first known (lower) temperature
- Read the Ambient Temperature of the Logger using the Realtime Screen
- Enter the first known temperature MINUS THE AMBIENT TEMPERATURE PLUS 32 °F in to the box labeled First Point Actual Value
- Click the First Point Button
- Allow the J thermocouple sensor to stabilize at the second known temperature
- Read the Ambient Temperature of the Logger using the Realtime Screen
- Enter the second known temperature MINUS THE AMBIENT TEMPERATURE PLUS 32 °F in to the box labeled Second Point Actual Value
- Click the Second Point Button
- The Calibration Offset and Calibration Gain figures will now be calculated and shown on the calibration screen.
- These Offset and Gain figures should now be written to the logger
- Click the Write Unit, Scaling and Calibration Values to logger button
- The Calibration process is now complete
If this procedure is used the J Thermocouple will show the correct temperature when sitting in free air.
If the ambient temperature is not subtracted from the known temperatures the J thermocouple will read high by the amount of the ambient temperature.