Light / Solar Energy Calibration
Light Calibration
- Place the Light Probe in Darkness
- Enter 0 into the First Point Actual Value box
- Click on the First Point button
- Place the Light Probe in bright sunlight next to a reference calibrated Light meter.
- Enter the light intensity value from the reference calibrated Light meter into the Second Point Actual Value box
- Click on the Second Point button
- Click on the Write Unit, Scaling and Calibration Values to logger button
The Calibration process is now complete
Solar Energy Calibration
- Place the Solar Energy Probe in Darkness
- Enter 0 into the First Point Actual Value box
- Click on the First Point button
- Place the Solar Energy Probe in bright sunlight next to a reference calibrated Solar Energy meter.
- Enter the Solar Energy value from the reference calibrated Solar Energy meter into the Second Point Actual Value box
- Click on the Second Point button
- Click on the Write Unit, Scaling and Calibration Values to logger button
The Calibration process is now complete